In the CMCs uncertainty statements, the notation Q[a, b] stands for the root-sum-square of the terms between brackets: Q[a, b] = [a2 + b2]1/2
 
Result of the search
Your selection : Thermometry
 
Slovakia, SMU (Slovensky Metrologicky Ustav)
Complete CMCs in Thermometry for Slovakia (.PDF file)
Humidity. Dew-point hygrometer, -60 °C to -45 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in °C: 0.3
Direct comparison
Approved on 03 November 2009
Internal NMI service identifier: SMU/2h
 
Humidity. Dew-point hygrometer, -45 °C to 20 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in °C: 0.2
Direct comparison
Approved on 03 November 2009
Internal NMI service identifier: SMU/3h
 
Temperature. Aluminium point, 660.323 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in mK: 2
By comparison
Temperature controlled furnace
Approved by 18 May 2004
Internal NMI service identifier: SMU/7
 
Temperature. Gallium point, 29.7646 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in mK: 0.22
By comparison
Temperature controlled thermostat
Approved by 18 May 2004
Internal NMI service identifier: SMU/3
 
Temperature. Indium point cell, 156.5985 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in mK: 0.95
Comparison with reference fixed-point cell
Temperature controlled furnace: three zone furnace
Approved on 20 May 2010
Internal NMI service identifier: SMU/4
 
Temperature. IPRT, -39 °C to 10 °C
Absolute expanded uncertainty (k = 2, level of confidence 95%) in °C: 0.01
Comparison with reference thermometer
Temperature controlled bath: stirred alcohol bath
Uncertainty at the calibration point. Hysteresis uncertainty for each IPRT must be added to the combined uncertainty quoted in the calibration report
Approved on 20 January 2010
Internal NMI service identifier: SMU/25


 