| In the CMCs uncertainty statements, the notation Q[a, b] stands for the root-sum-square of the terms between brackets: Q[a, b] = [a2 + b2]1/2 |
| |
Result of the search |
 |
 |
Your selection : Ionizing Radiation, Neutron Measurements, Brazil
|
| | | Brazil, LNMRI (Laboratorio Nacional de Metrologia das Radiaçoes Ionizantes) | | Complete CMCs in Ionizing Radiation for Brazil (.pdf file) |  | | Ambient dose equivalent rate. Neutron dosemeter, 1.2E-04 Sv h-1 to 3.7E-04 Sv h-1 | | Relative expanded uncertainty (k = 2, level of confidence ~95%) in %: 7.0 | | Irradiation with a calibrated neutron source | | Neutron spectrum at 1 m from the source: Am-241/Be-9 ISO 8529-3 | | Reference standard: Calibrated Am-241/Be-9 source (SN 4 080) | | Source of traceability: LNE | | Approved on 21 October 2005 | | Internal NMI service identifier: SIM-RAD-LNMRI-3002 |
| | | Emission rate. Sealed neutron source, 1.5E+05 s-1 to 1.0E+07 s-1 | | Relative expanded uncertainty (k = 2, level of confidence ~95%) in %: 2.3 | | Measurement in a manganese bath | | Am-241/Be-9 source: no further specifications | | Reference standard: Standardized Mn-56 | | Source of traceability: LNMRI | | Approved on 21 October 2005 | | Internal NMI service identifier: SIM-RAD-LNMRI-3001 |
|