Calibration and Measurement Capabilities Photometry and Radiometry |
 |
| In the CMCs uncertainty statements, the notation Q[a, b] stands for the root-sum-square of the terms between brackets: Q[a, b] = [a2 + b2]1/2 |
| |
Result of the search |
 |
Your selection : Photometry and Radiometry
|
| | | Czech Republic, CMI (Czech Metrology Institute) |
| | Complete CMCs in Photometry and Radiometry for Czech Republic (.PDF file) |  | | Luminous flux. Tungsten lamp, 10 lm to 10000 lm | | Relative expanded uncertainty (k = 2, level of confidence 95%) in %: 2.1 | | Integrating sphere, substitution method | | Correlated color temperature: 2600 K to 3100 K | | Approved on 30 September 2010 |
| | | Luminous intensity. Tungsten lamp, 1 cd to 10000 cd | | Relative expanded uncertainty (k = 2, level of confidence 95%) in %: 2 | | Photometric bench and reference lamps / photometer | | Correlated colour temperature: 2000 K to 3200 K | | Approved on 30 September 2010 |
| | | Responsivity laser, power. Si-Trap detectors. Measurand unit, A/W or V/W | | Relative expanded uncertainty (k = 2, level of confidence 95%) in %: 0.05 | | Cryogenic radiometer and laser | | Wavelength range: laser lines from 356 nm to 413 nm | | Power level: < 1 mW | | Other types of detector can also be calibrated Approved on 30 September 2010 |
| | | Responsivity laser, power. Si-Trap detectors. Measurand unit, A/W or V/W | | Relative expanded uncertainty (k = 2, level of confidence 95%) in %: 0.03 | | Cryogenic radiometer and laser | | Wavelength range: laser lines from 476 nm to 799 nm | | Power level: < 1 mW | | Other types of detector can also be calibrated Approved on 30 September 2010 |
| | | Responsivity spectral, power. Broad band detectors. Measurand unit, A/W or V/W | | Relative expanded uncertainty (k = 2, level of confidence 95%) in %: 0.95 | | Double grating monochromator | | Wavelength range: 300 nm to 405 nm | | Bandwidth: < 10 nm | | Power level: < 30 µW | | Other types of detector can also be calibrated Approved on 30 September 2010 |
| | | Responsivity spectral, power. Broad band detectors. Measurand unit, A/W or V/W | | Relative expanded uncertainty (k = 2, level of confidence 95%) in %: 0.4 | | Double grating monochromator | | Wavelength range: 405 nm to 920 nm | | Bandwidth: < 10 nm | | Power level: < 30 µW | | Other types of detector can also be calibrated Approved on 30 September 2010 |
|