Home    Key and supplementary comparisons    Calibration and Measurement Capabilities - CMCs
CMCs - Result of the search
Physics
Acoustics, Ultrasound, Vibration
Electricity and Magnetism
Length
Mass and related quantities
Photometry and Radiometry
Ionizing Radiation
Thermometry
Time and Frequency
Chemistry
Chemistry
Related links
KCDB Statistics
KCDB FAQs
KCDB Reports
CIPM MRA
JCRB
Find my NMI
Metrologia
Contact us
BIPM.KCDB@bipm.org
Calibration and Measurement Capabilities
Acoustics, Ultrasound, Vibration
In the CMCs uncertainty statements, the notation Q[a, b] stands for the root-sum-square of the terms between brackets: Q[a, b] = [a2 + b2]1/2
 
Result of the search
Your selection : Acoustics, Ultrasound, Vibration
 
Korea, Republic of, KRISS (Korea Research Institute of Standards and Science)
Complete CMCs in Acoustics, Ultrasound, Vibration for Korea, Republic of (.PDF file)
Acceleration (modulus). Acceleration calibrator, 10 m/s2 to 100 m/s2
Relative expanded uncertainty (k = 2, level of confidence 95%) in %: 2
ISO 16063-21:2003 (comparison)
Frequency: 80 Hz or 160 Hz
Approved on 26 April 2017
Internal NMI service identifier: KRISS/C63211
 
Acceleration (modulus). Acceleration measuring instrument (with accelerometer), 10 m/s2 to 100 m/s2
Relative expanded uncertainty (k = 2, level of confidence 95%) in %: 0.5
ISO 16063-11:1999 (interferometry)
Frequency: 20 Hz to 250 Hz
Approved on 26 April 2017
Internal NMI service identifier: KRISS/C63111
 
Acceleration (modulus). Acceleration measuring instrument (with accelerometer), 10 m/s2 to 200 m/s2
Relative expanded uncertainty (k = 2, level of confidence 95%) in %: 1
ISO 16063-11:1999 (interferometry)
Frequency: > 250 Hz to 5 kHz
Approved on 26 April 2017
Internal NMI service identifier: KRISS/C63111
 
Angular acceleration (modulus). Angular acceleration measuring instrument (with angular accelerometer), 10 rad/s2 to 100 rad/s2
Relative expanded uncertainty (k = 2, level of confidence 95%) in %: 0.5
ISO 16063-15: 2006 (interferometry)
Frequency: 1 Hz to 8 Hz
Approved on 26 April 2017
Internal NMI service identifier: KRISS/C64111
 
Angular acceleration (modulus). Angular acceleration measuring instrument (with angular accelerometer), 50 rad/s2 to 200 rad/s2
Relative expanded uncertainty (k = 2, level of confidence 95%) in %: 0.3
ISO 16063-15: 2006 (interferometry)
Frequency: 10 Hz to 160 Hz
Approved on 26 April 2017
Internal NMI service identifier: KRISS/C64111
 
Angular acceleration (modulus). Angular acceleration measuring instrument (with angular accelerometer), 50 rad/s2 to 200 rad/s2
Relative expanded uncertainty (k = 2, level of confidence 95%) in %: 0.8
ISO 16063-15: 2006 (interferometry)
Frequency: 200 Hz to 800 Hz
Approved on 26 April 2017
Internal NMI service identifier: KRISS/C64111


 